The idea of applying mass spectrometry to analyzing dissolved analyte species in liquids originates from 1963 where Hoch and Kok introduced a technique called membrane inlet mass spectrometry (MIMS).1 In MIMS, volatile species permeate through semi-permeable polymer membranes, typically polydimethylsiloxane (PDMS), directly into a mass spectrometer. MIMS systems are capable of highly sensitive measurements, especially if the membrane material is chosen for preferential permeation of the analyte species of interest, which causes an up-concentration of analyte inside the membrane.

 

MIMS systems, however, suffer from a slow time-response due to the diffusive transport mechanism through the membrane material, which makes them unsuitable for timeresolved electrochemistry measurements. In 1971 Bruckenstein and Gadde improved membrane transport drastically by changing to a perforated and hydrophobic membrane made of porous polytetrafluoroethylene (PTFE), through which both water and analyte could evaporate much quicker.2

Schematic representation of the unique Spectro Inlets EC-MS microchip-based inlet
 Hydrophobic membrane inlet from Bruckenstein and Gadde

These first cell congurations were all able to capture 100% of the analyte produced at the working electrode given that the electrode was smaller than the membrane area and placed in close proximity to the membrane, as long as the faradaic current was kept small (1 mA/cm2). This introduces the concept of a membrane collection efficiency, relating the amount of analyte reaching the vacuum system, to the amount of analyte produced at the working electrode. The response time in the stagnant thin-layer cell is fast but limited by the diffusion across the electrolyte layer. For less volatile species, it can furthermore be limited by evaporation across the membrane interface, causing both time delay and up-concentration of analyte species in the thin-layer working volume.

Both diffusion and evaporation vary between different analyte species. Furthermore, the stagnant thin-layer cell does not allow for the use of pre-purged electrolytes as the working volume gets, by design, immediately depleted from purging gas. In order to lock down the time-response and allow the use of pre-purged electrolytes, a dual thin-layer flow cell was introduced by Jusys, et al. in 1999,4 which utilized a thin-layer working volume and a corresponding thin-layer collection volume placed above a porous PTFE membrane downstream from the working electrode. These cells typically have an internal volume of ~5 μl and a continuous electrolyte flow of ~5 μl/s, giving a time response of  1 s.5

 

MIMS system from Hoch, G. & Kok, B.

Since 1984, DEMS has undergone a huge development with the focus being on electrochemistry cell design. The first “classical” DEMS cell designs from 1971 and 1984 utilized deposition of electrode material directly onto the membrane inlet, allowing for a fast transport of analyte within the electrolyte.

In 1993 Baltruschat introduced a stagnant thinlayer cell,3  allowing for the use of massive working electrodes placed directly opposite of the membrane inlet with a 100 μm thick Teflon (PTFE) spacer used to define the working distance between the electrode and the membrane, forming a thinlayer electrochemistry working volume between the two. 

Schematic representation of the unique Spectro Inlets EC-MS microchip-based inlet
Stagnant thin-layer cell from Baltruschat et al.

However, at these flow-rates 100% membrane collection efficiency is no longer possible, especially for less volatile species, thus compromising the overall sensitivity.

Throughout the years many other cell designs have been introduced, including hanging meniscus configurations, which are particularly useful for single crystal studies,6–8 and other more exotic cell designs like rotating disk electrode and scanning flow cell configurations.9–1

 

Schematic representation of the unique Spectro Inlets EC-MS microchip-based inlet
Dual thin-layer flow cell from Juzys et al
Schematic representation of the unique Spectro Inlets EC-MS microchip-based inlet
DEMS cell as developed by Baltrushat and coworkers. J. Am. Soc. Mass Spectrom.
15, 1693–1706 (2004).

DEMS has also been coupled to complementary detection techniques like e.g. infrared spectroscopy (IR)12 and electrochemical detection methods.13,14

Most of the above-mentioned DEMS configurations utilize similar membrane inlet systems constituting of nanoporous PTFE membranes and differential pumping.

Other inlet systems have been introduced, which circumvent the need for differential pumping by decreasing the inlet area and thus probing a single location on an electrode, typically in a hanging meniscus configuration.7

These systems are often referred to as on-line electrochemistry mass spectrometry (OLEMS) systems. OLEMS cell configurations tend to be more versatile to use but are incapable of quantitative analysis.

OLEMS should not be confused with system configurations sampling the head space (gas volume) above the electrolyte surface in a larger electrochemistry cell, often referred to as OEMS.15

These systems do also perform “on-line” analysis but calling it real-time would be a misinterpretation. Since its introduction, DEMS has been widely used to gain insight into electrochemical reaction mechanics, with ethanol 16 and methanol17 oxidation, complex electrocatalytic systems like electrochemical CO2 hydrogenation18 and fundamental battery research19  representing a few examples.

The above is only a selection of key developments in the history of membrane inlet mass spectrometry. A detailed review of the historical development of MIMS can be found in,20 reviews on specific polymer membrane MIMS systems (oriented towards biochemistry) can be found in 20,21,30,22–29 and reviews on conventional DEMS systems (for electrochemistry) can be found in.5,31,32 

Schematic representation of the unique Spectro Inlets EC-MS microchip-based inlet
Hanging meniscus cell from Gao et al.
Schematic representation of the unique Spectro Inlets EC-MS microchip-based inlet
Hanging meniscus cell for single crystal studies,
from Wonders et al.

Schematic representation of the unique Spectro Inlets EC-MS microchip-based inlet
Rotating disk cell from Tegtmeyer et al.
Schematic representation of the unique Spectro Inlets EC-MS microchip-based inlet
OEMS cell from Tsiouvaras et al.