EC‐MS Technical Note #18

Ex‐situ gas injection cell

EC-MS Technical notes #18

Introduction

The gas injection cell provides an interface for introducing gas samples into the EC‐MS ex‐situ, for example, a gas sample extracted from a battery pouch cell. The cell consists of a body with a slot for a septum and a screw cap with a needle injection port ﴾see fig 1a﴿. When the cell is assembled, the screw cap holds and compresses the septum to seal off the cell ﴾see fig. 1b﴿, which allows for controlled gas injections into the EC‐MS.

The gas injection cell is fastened to the EC‐MS interface with four M3 x 6 bolts tightened to 10 N cm, and also uses an O‐ring to seal between the bottom face of the cell body and the chip, like other EC-MS cells. Note, that any chip ﴾standard or non‐aqueous﴿ may be used, but the cell is only intended for gas injection ‐ no liquids. Since standard chips are generally more robust, these are recommended for this use. To protect the chip, the mass spectrometer and vacuum system components, no volume of gas should be forced into the cell abruptly. The pressure of the gas in the syringe ﴾prior to injection﴿ should be near ambient ﴾≈ 1 bar﴿. Specifications for suitable septa and syringes ﴾including needles﴿ are given in table 1.